JJF(黔)59-2021
微量进样器校准规范
Cal ibration Specification of Microsampling Syringe
目 录
引言............................................................. (1)
1 范围............................................................ (1)
2 引用文件........................................................ (1)
3 术语............................................................ (1)
4 概述............................................................ (2)
5 计量特性........................................................ (2)
6 校准条件........................................................ (3)
6.1 环境条件 ...................................................... (3)
6.2 校准介质 ...................................................... (3)
6.3 校准设备 ...................................................... (3)
7 校准项目和校准方法 .............................................. (4)
7.1 校准项目 ...................................................... (4)
7.2 校准方法....................................................... (4)
8 校准结果的表达.................................................. (6)
9 复校时间间隔.................................................... (7)
附录 A 衡量法 K(t)值表 ........................................... (8)
附录 B 微量进样器校准值的测量不确定度评定示例..................... (10)
附录 C 微量进样器校准记录......................................... (15)
附录 D 校准证书(内页)格式....................................... (16)
引 言
本规范依据 JJF 1071-2010《国家计量校准规范编写规则》、JJF 1001-2011《通用计量术语及定义》、JJF 1059.1-2012《测量不确定度评定与表示》编制。
1 范围
本规范适用于测量范围为(
1~1000)μL 的微量进样器(以下简称进样器)的校准。
2 引用文件
本规范引用了下列文件:
JJG 196-2006 常用玻璃量器
JJG 646-2006 移液器
GB/T 6682-2008 分析实验用水规格和试验方法
YY 0088-1992 微量进样器
凡是注日期的引用文件,仅注日期的版本适用于本规范;凡是不注日期的引用文件,
其最新版本(包括所有的修改单)适用于本规范。